| Cardiomedical | Products Cardio Vision | Pacing/Heartwire

Pacing

External Pacemaker Pacing programmable parameters:

Advantages:

External Pacemaker Pacing Programmable Parameters:
Normal Rate: 30-200ppm |step:5ppm| Accuracy:+/-5%
Pulse Amplitude: 0.2V-10V | step:0.2V | Accuracy:+/-5%
Sensitivity: 0.2mV-12mV | step: 0.2mV | Accuracy:+/-0.2mV
Refractory period: 200ms-400ms | step: 5ms | Accuracy:+/-1.5%
Pulse Width: 01ms-2ms | step: 0.1ms | Accuracy: +/-0.2%
Hysteresis: 0 - 5 - 10 - 15 - 20ppm
Emergency: SSI(AAI/VVI) | R=70ppm | A=10V | S=2.4mV | W=1ms
EMI Protection: auto-switching to AOO/ VOO with acoustic alarm
Defibrillation Protection: Up to 400Joules
Software Failure Protection: auto-switching to Emergency mode
High-rate Operation (overdrive): 120ppm-600ppm | step: 10ppm | Accuracy:+/-3%. Patient heart-rate(HR) is measured automatically and overdrive rate is set 30% higher. For safety reasons, Galix external pacemaker generates auto-quit if no key is pressed within a 20s period.

Measurements and Statistics: R-R/P-P Interval: 5ms-2s
Patient Frequency: 30bpm to 300bpm
P-Wave or R-Wave Amplitude Pacing clock: Up to 10 days | Accuracy:+/-0.01%
Pacing Percentage: ratio between the total number of paced beats, and total number of paced beats plus spontaneous beats
Low Battery Indicator: Acoustic and visual alarms are activated once batteries can not withstand 3 hours of operation

External Pacemaker, Temporary Myocardial Heartwires

External Pacemaker, Temporary Myocardial Heartwires

CV10-Single
CV20-Dual
Temporary Myocardial Heartwires

 

 

 

Download

cataloge PDF downloadcataloge PDF download (22 MB)

 



                                                           
      © 2009  